![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 354](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996354.webp)
Agilent Technologies
Using a Swagelok Tee
To supply gas from a single source to more than one input,
use a Swagelok Tee.
Materials needed:
•
1/8- inch preconditioned copper tubing
•
Tubing cutter
•
1/8- inch Swagelok nuts and front and back ferrules
•
1/8- inch Swagelok Tee
•
Two 7/16- inch wrenches
•
1/8- inch Swagelok cap (optional)
1
Cut the tubing where you want to install the Tee. Connect
the tubing and Tee with a Swagelok fitting. See
Figure 15
Swagelok tee
2
Measure the distance from the Tee to the GC inlets.
Attach copper tubing to the open Tee ends with Swagelok
fittings.
N O T E
Do not combine valve actuator air with flame ionization air. The valve
action will cause major upsets in the detector signal.
Cap
Tee fitting
Nut and ferrules
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...