![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 318](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996318.webp)
318
Maintaining Your GC
16
Maintaining the G3435A/G3436A FPD
To Remove the FPD Vent Tube
1
Gather the following:
•
T- 20 Torx screwdriver
•
9/16- inch wrench
2
Load the
and wait for the GC to
become ready.
3
Drain any water from the flexible vent tubing and
discard.
4
Open the FPD cover.
5
Remove the flexible tubing from the FPD vent tube.
6
Loosen and remove the vent tube assembly with a
wrench.
7
Remove the screws securing the FPD cover.
•
The single- wavelength detector has two screws at the
bottom of the left side (top photo below) and two
screws at the top of the right side (bottom photo
below).
•
The dual- wavelength detector has two screws at the
top of the right side (bottom photo below).
C A U T I O N
When turning off the GC, first turn off the flame to prevent
condensation from dripping into the jet and column.
WA R N I N G
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...