![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 112](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996112.webp)
112
Maintaining Your GC
7
Maintaining the COC Inlet
To Change a Septum on the COC Inlet
1
Gather the following:
•
Replacement septum. (See
•
Tweezers
•
A thin wire (0.2- inch diameter) for removing septum
from inlet
•
Lint- free gloves
2
Load the
and wait for the GC
to become ready.
3
Replace the septum.
•
If you are using a septum nut, grasp the knurling and
unscrew. Remove the old septum with tweezers. Use
tweezers to install a new septum. Push the septum into
the septum nut until properly seated. Firmly tighten
the nut.
WA R N I N G
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands
.
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
For 250/320-µm
automated injections
For 530-µm automated
injections
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...