![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 145](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996145.webp)
Maintaining the MMI
8
Maintaining Your GC
145
5
Completely loosen the two screws that secure the split
vent valve in place.
6
Lift the filter trap assembly and split vent valve from the
mounting bracket together and unscrew the split vent
front weldment on the filter trap assembly. Be careful not
to stress the tubing between the split vent valve and the
trap.
7
Remove the old filter cartridge and two O- rings.
8
Verify the new O- rings are seated properly on the new
filter cartridge.
9
Install the new filter cartridge then reassemble the trap.
Do not fully tighten yet.
10
Place the filter trap assembly in the mounting bracket and
install the retaining clip.
11
Install the split vent valve.
12
Fully tighten the split vent front weldment onto the trap.
13
Check for leaks.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...