Maintaining the GC
3
Maintaining Your GC
25
To Change the Hydrogen Sensor Calibration Gas Cylinder
Replace the calibration gas cylinder when you cannot adjust it
to provide the correct flow rate during calibration, or if it
expires.
1
Make sure the old cylinder is empty.
2
Turn the pressure regulator completely off and set the
output pressure as low as possible (full CCW).
3
Remove the pressure regulator from the old cylinder and
install it onto the new cylinder.
4
Install the new cylinder into the stand.
5
Turn on the pressure in the new cylinder.
6
Open the GC oven door and connect a flowmeter tube to the
sensor tube in the oven.
7
Press
[Options]
then go to
Calibration
>
Hydrogen Sensor
and
press [
Enter
].
8
Scroll to
Start Calibration Cycle?
and press [
On/Yes
]. The
calibration cycle begins. The hydrogen sensor module will
wait to stabilize, then will begin to send calibration gas
across the sensor.
WA R N I N G
Connect only the Agilent calibration gas to the hydrogen sensor
fitting. The calibration gas is 2% hydrogen in air and is not
potentially explosive. Gases with higher concentrations of
hydrogen can create a dangerous condition in the oven or damage
the hydrogen sensor.
&RQQHFWWKHÁRZ
meter to the
hydrogen sensor
tube next to the
oven heater shroud
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...