Maintaining the FID
11
Maintaining Your GC
207
If the column id is greater than 0.1 mm:
a
Gently insert the column into the detector until it
bottoms; do not attempt to force it further.
b
Finger- tighten the column nut, then withdraw the
column about 1 mm. Tighten the nut an additional
1/4 turn with a wrench.
If the column id is 0.1 mm or less position the column so
it extends above the ferrule by 48 mm (
capillary
optimized
fitting) or 68 mm (
adaptable
fitting). Slide the
septum up to hold the column nut and ferrule at this fixed
position.
c
Insert the column into the detector. Slide the nut and
ferrule up the column to the detector base.
Finger- tighten the column nut until it grips the
column.
d
Adjust the column (
not
the septum) position so that
the septum is even with the bottom of the column nut.
Tighten the nut an additional 1/4 turn with a wrench.
48 mm
68 mm
Capillary optimized fitting
Adaptable fitting
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...