![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 212](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996212.webp)
212
Maintaining Your GC
11
Maintaining the FID
To Replace an FID Jet
1
Gather the following:
•
Replacement jet (See
•
T- 20 Torx screwdriver
•
1/4- inch nut driver
•
Tweezers
•
Compressed, filtered, dry air or nitrogen
•
Solvent that will clean the type of deposits in your
detector
•
Clean cloth
•
Cotton swab
•
Lint- free gloves
2
Load the
and wait for the GC to
become ready.
3
If installed, remove the capillary column from the
detector.
4
Remove the FID collector assembly and place it on a
clean cloth. (See
“To Replace the FID Collector Assembly”
WA R N I N G
Be careful! The oven and/or detector may be hot enough to cause
burns. If the detector is hot, wear gloves to protect your hands.
WA R N I N G
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...