Maintaining the NPD
14
Maintaining Your GC
281
9
Remove the large metal C- ring and the upper ceramic
insulator if they were not attached to the lid.
10
Remove the collector. If the detector is operated at high
temperatures, the collector parts may stick inside the
detector. Gently push and wiggle them to break the seal.
11
Use tweezers to remove the lower ceramic insulator and
the two small metal C- rings located above and below the
collector. If these parts are stuck together, do not
separate them. If they are not stuck, remember which
metal ring was on top of the insulator and which was
below it. The pieces must be reassembled in the same
orientation.
12
If not replacing the jet, skip to
13
Remove the column from the detector.
Figure 7
NPD collector, upper insulator, and metal C-ring
Figure 8
NPD lower ceramic insulator and metal C-rings
Metal C-ring, top
Ceramic insulator, upper
Collector
Metal C-ring, bottom
Ceramic insulator, lower
Metal C-ring, bottom
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...