![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 118](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996118.webp)
118
Maintaining Your GC
7
Maintaining the COC Inlet
To Replace the 7693A Injector Needle Support Guide
Before using a 7693A ALS to make injections onto the COC
inlet, install the on- column needle support guide.
Refer to the
for complete
details.
1
Open the injector door.
2
Remove the syringe.
3
Slide the syringe carriage up to the top position.
4
Completely remove the T- 10 Torx screw from the support
foot. Be careful to not let the screw fall into the turret
assembly.
5
Slide off the support foot.
6
Slide on the new support foot.
7
Replace the T- 10 Torx screw and tighten.
8
Install the appropriate syringe.
9
Close the injector door.
10
Align the injector.
Standard needle
support guide
On-column needle
support guide
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...