![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 48](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996048.webp)
48
Maintaining Your GC
5
Maintaining the Split/Splitless Inlet
To Install a Capillary Column with the Split/Splitless Inlet
1
Gather the following (see
“Consumables and Parts for the
•
Column
•
Ferrule(s)
•
Column nut
•
Septum
•
Column cutter
•
Isopropanol
•
Lab tissue
•
Metric ruler
•
1/4- inch open- end wrench
•
Lint- free gloves
2
Load the
and wait for the GC to
become ready.
3
Verify that the correct glass liner is installed. (See
“Consumables and Parts for the Split/Splitless Inlet”
4
Place the column on the hanger with the ends pointing up
and the label to the front.
WA R N I N G
Do not use hydrogen as the carrier for conditioning! It could vent
into the oven and present an explosion hazard.
WA R N I N G
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands
.
WA R N I N G
Wear safety glasses to protect your eyes from flying particles
while handling, cutting, or installing glass or fused silica capillary
columns. Use care in handling these columns to prevent puncture
wounds.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...