14
Maintaining Your GC
1
About Maintaining the GC
•
The filament in the TCD will be damaged if exposed to
air while hot. To protect the filament, turn it
Off
.
•
Set all detector flows to
Off
.
Once the zones reach < 70 °C, you can perform general GC
maintenance.
Inlet Maintenance Method
This method prepares the inlet for maintenance while
leaving the detector at operating temperature.
•
Set the oven temperature to
35 °C
. This allows the oven
fan to assist cooling.
•
Set all inlet temperatures to
Off
and set inlet gas
pressures to
0.0
.
•
If performing column maintenance, remember to wait
for the oven and column to cool down before turning
off column carrier gas flow at the source. Also
remember to cap both ends of the column to keep air
out once it is removed.
•
If you are not performing column maintenance, keep
inert carrier gas (helium or nitrogen) flowing to protect
the column.
•
Maintain all temperature setpoints for installed detectors,
if desired.
•
The filament in the TCD will be damaged if exposed to
air while hot. To protect the filament, turn it
Off
.
Once the zones reach < 70 °C, you can perform general GC
maintenance.
WA R N I N G
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
WA R N I N G
Be careful! The oven, inlet, and/or detector may be hot enough to
cause burns. If the oven, inlet, or detector is hot, wear
heat-resistant gloves to protect your hands.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...