Maintaining the PTV Inlet
9
Maintaining Your GC
169
11
Insert the glass liner into the inlet from above until the
unpacked side of the ferrule rests on the top of the inlet.
12
Replace the head:
•
For septumless head, screw the head onto the inlet and
tighten 1/8 turn past finger- tight with a wrench.
Reconnect the carrier gas line.
•
For septum head, align the head with the inlet and
manually engage the free- spinning nut to the inlet.
Tighten 1/2 turn past finger- tight with a wrench.
13
Check all connections for leaks. If necessary, tighten them
again by hand.
14
Restore the analytical method.
15
Configure the new liner. If using the optional barcode
scanner, scan the liner directly into your data system
software configuration, then download the changes to the
GC.
16
Reset the liner counters. See
in
the
Operation Manual
.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...