Maintaining the MMI
8
Maintaining Your GC
143
Clean the inlet if there is visible or suspected
contamination. (See
“To Clean the Multimode Inlet”
Clean O- ring residue from sealing surface.
7
Slide a new O- ring onto the replacement liner.
8
Return the liner to the inlet, pushing it all the way in
until the liner contacts the seal.
9
Line up the tab on the bottom of the septum assembly
with the slot on the insert assembly and push down to
connect. Slide the locking tab to the back.
10
Turn on the inlet. Allow the inlet and column to purge
with carrier gas for 15 minutes before heating the inlet or
the column oven.
11
Bakeout contaminants. (See
12
Restore the analytical method.
13
Reset the liner counter. See
the
Operation Manual
.
14
If using the optional barcode scanner, scan the liner
information directly into your data system software
configuration, then download the changes to the GC.
15
Check for leaks.
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...