![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 92](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996092.webp)
92
Maintaining Your GC
6
Maintaining the Purged Packed Inlet
To Install a Packed Column Adapter on a Detector Fitting
1
Gather the following:
•
7/16- inch, 9/16- inch, and 1/2- inch wrenches
•
Vespel/graphite ferrule (See
•
Brass column nut
•
Lint- free gloves
•
Adapter.
Select the appropriate adapter from one of the parts lists
shown below:
•
Consumables and Parts for the FID
(Packed columns
can only be installed on an adaptable FID.)
•
Consumables and Parts for the TCD
Consumables and Parts for the NPD
Consumables and Parts for the FPD
•
For uECD, 1/4- inch packed columns require no
adapter. For 1/8- inch packed columns, see
Consumables and Parts for the uECD
2
Load the
and wait for the GC to
become ready.
3
Assemble a nut and a ferrule onto the adapter.
WA R N I N G
Be careful! The detector may be hot enough to cause burns. If the
detector is hot, wear heat-resistant gloves to protect your hands.
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...