![Agilent Technologies Agilent 7890 Series Maintenance Manual Download Page 99](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996099.webp)
Maintaining the Purged Packed Inlet
6
Maintaining Your GC
99
To Install Ferrules on a Packed Metal Column
1
Gather the following:
•
Wrenches
•
Stainless steel male Swagelok fitting, 1/4- or 1/8- inch
od
•
Brass Swagelok nut and ferrule set (See
and Parts for the Purged Packed Inlet”
•
Lint- free gloves
2
Verify that the column end is cut square and is free of
burns and deformation.
3
Secure the fitting in a bench vise.
4
Assemble a Swagelok nut and ferrules onto the column.
5
Fully insert the column into the vise- held fitting, then
withdraw 1–2 mm. Finger- tighten the nut.
6
Tighten the nut an additional 3/4 turn with a wrench (for
a 1/8- inch column) or an additional 1- 1/4 turn (for a
1/4- inch column).
7
Unscrew the column nut from the vise- held fitting and
remove the column. Ferrules should now be set in place
on the column with the column end correctly positioned.
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Front ferrule
Back ferrule
Column nut
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...