Maintaining the VI
10
Maintaining Your GC
183
8
Position the column so it extends 6 mm above the end of
the ferrule. Slide the septum up the column to hold the
column nut at this fixed position.
9
Insert the column into the interface and finger- tighten the
column nut.
10
Adjust the column (
not
the septum) position until the
septum is snug against the bottom of the nut.
11
Tighten the column nut an additional 1/4 to 1/2 turn with
a wrench so that the column cannot be pulled from the
fitting with gentle pressure.
12
Configure the new column. If using the optional barcode
scanner, scan the column directly into your data system
software configuration, then download the changes to the
GC.
13
With the sample transfer line attached and the column
attached to the inlet and detector, establish a flow of
carrier gas through the transfer line. Purge as
recommended by the column manufacturer.
14
Condition the column per the manufacturer’s
recommendation. (See
To Condition a Capillary Column
.)
15
Install the column into the detector. See:
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
•
To Install a Capillary Column in the uECD
To install a Capillary Column in the FPD Plus
To Install a Capillary Column Adapter in the FPD
16
After the column is installed at both inlet and detector,
establish a flow of carrier gas and purge as recommended
by the column manufacturer.
17
Restore the analytical method.
•
For FPD, immediately turn off the flame.
•
For NPD, immediately set the bead voltage to 0.0.
18
After the GC becomes ready, wait 10 minutes then ignite
the detector flame or adjust offset on the NPD bead.
6 mm
Summary of Contents for Agilent 7890 Series
Page 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Page 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Page 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Page 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Page 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Page 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Page 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Page 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Page 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Page 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Page 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Page 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Page 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Page 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Page 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Page 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Page 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Page 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...