Disassembly
Main Component Disassembly
1. Remove processor board. Turn instrument over. Remove the T-8 screw securing the processor board. Press the
tabs on the processor board connector and slide processor board toward the back of instrument to disengage the
connector. Lift processor board out.
2. Remove front panel assembly. Remove T15 screw holding the main board. Press latch on left side of front panel
and latch in power supply cover on right side of front panel. Push sides of metal chassis toward center to disengage
studs on sides of front panel assembly. Gently pull front panel assembly straight off chassis. Note that front panel
Agilent 33500 Series Operating and Service Guide