ABORt
ABORt
Halts a sequence, list, sweep, or burst, even an infinite burst. Also causes trigger subsystem to return to idle state. If
is ON, instrument immediately proceeds to wait-for-trigger state.
Parameter
Typical Return
(none)
(none)
Halt the items listed above:
ABOR
l
Halts any triggered action (triggered list, triggered sweep, triggered burst, triggered arbitrary waveform playback).
l
ABORt has no effect when instrument is in normal or modulated modes, except for sequenced arbitrary wave-
forms, lists, bursts, and sweeps. If instrument is running a sequence, list, burst, or sweep, ABORt restarts the
stopped item with the current
and trigger conditions.
l
When ABORt occurs in list mode, the frequency goes back to the "normal" mode frequency until the first trigger
occurs. After the first trigger, the first frequency in the list will be used.
l
If ABORt executed during sweep, sweep returns to starting sweep frequency.
l
ABORt always applies to both channels in a two-channel instrument.
156
Agilent 33500 Series Operating and Service Guide