APPLy Subsystem
Commands and Queries
[SOURce[1|2]:]APPLy?
Queries the output configuration.
Parameter
Typical Return
(none)
"SIN +5.00000000003,
+3.000000000,
-2.500000000"
Return the configuration for a 5 kHz, 3 V sine wave with a -2.5 VDC offset.
APPLY?
l
The function, frequency, amplitude, and offset are returned as shown above. The amplitude, but not the offset, is
returned as specified by
.
[SOURce[1|2]:]APPLy:ARBitrary [{<
sample_rate
>|MIN|MAX|DEF}
[,{<
amplitude
>|MIN|MAX|DEF} [,{<
offset
>|MIN|MAX|DEF}]]]
Outputs arbitrary waveform selected by FUNCtion: ARBitrary, using the specified sample rate, amplitude, and offset.
Parameter
Typical Return
<
sample_rate
> from 1 µSa/s to 250 MSa/s, default 40 kSa/s
(none)
<
amplitude
> from 1 mVpp to 10 Vpp into 50 Ω, 2 mVpp to 20 Vpp into an open circuit,
default 100 mVpp into 50 Ω
<
offset
> is the DC offset voltage (default 0), from ±5 VDC into 50 Ω, or from ±10 VDC
into an open circuit.
Output the arbitrary waveform selected using FUNCtion:ARBitrary:
APPLy:ARBitrary 1 KHZ, 5.0, -2.5 V
General
164
Agilent 33500 Series Operating and Service Guide