APPLy Subsystem
Parameter
Typical Return
<
amplitude
> Desired output amplitude in Vpp, Vrms or dBm, as specified by
.
1 mVpp to 10 Vpp into 50 Ω, or twice that into an open circuit. Default 100 mVpp into 50 Ω.
<
offset
> is the DC offset voltage (default 0), from ±5 VDC into 50 Ω, or from ±10 VDC into
an open circuit.
Output 3 Vpp sine wave at 5 kHz with -2.5 V offset.
APPL:SIN 5 KHZ, 3.0 VPP, -2.5 V
Offset Voltage
l
The relationship between offset voltage and output amplitude is shown below. Vmax is the maximum peak voltage
for the selected output termination (5 V for a 50 Ω load or 10 V for a high-impedance load).
|Voffset| < Vmax - Vpp/2
If the specified offset voltage is not valid, the instrument will adjust it to the maximum DC voltage allowed with the
specified amplitude. From the remote interface, a "Data out of range" error will also be generated.
l
Limits Due to Output Termination: The offset range depends on the output termination setting. For example, if
you set offset to 100 mVDC and then change output termination from 50 Ω to "high impedance," the offset volt-
age displayed on the front panel doubles to 200 mVDC (no error is generated). If you change from "high imped-
ance" to 50 Ω, the displayed offset voltage will be halved. See
for details.
[SOURce[1|2]:]APPLy:SQUare [{<
frequency
>|MIN|MAX|DEF}
[,{<
amplitude
>|MIN|MAX|DEF} [,{<
offset
>|MIN|MAX|DEF}]]]
Outputs a square wave with the specified frequency, amplitude, and DC offset. In addition,
performs the fol-
lowing operations:
l
Overrides the current duty cycle setting (FUNCtion:SQUare:DCYCle), and sets a 50% duty cycle for the square
wave.
Parameter
Typical Return
<
frequency
>
from 1 μHz to instrument's maximum frequency.
Default 1 kHz.
(none)
<
amplitude
> Desired output amplitude in Vpp, Vrms or dBm, as specified by
.
1 mVpp to 10 Vpp into 50 Ω, or twice that into an open circuit. Default 100 mVpp (into 50 Ω)
<
offset
> is the DC offset voltage (default 0), from ±5 VDC into 50 Ω, or from ±10 VDC into
an open circuit.
Output 3 V square wave at 5 kHz with -2.5 V offset:
APPL:SQU 5 KHZ, 3.0 V, -2.5 V
170
Agilent 33500 Series Operating and Service Guide