LAN Configuration
SYSTem:COMMunicate:LAN:SMASk "<
mask
>"
SYSTem:COMMunicate:LAN:SMASk? [{CURRent | STATic}]
Assigns a subnet mask for the instrument. The instrument uses the subnet mask to determine whether a client IP
address is on the same local subnet. When a client IP address is on a different subnet, all packets must be sent to the
Default Gateway. Contact your LAN administrator for details.
If you change this setting, you must send
to activate the new
setting.
Parameter
Typical Return
Command: "
", default "0.0.0.0"
Query: {CURRent|STATic}, default CURRent
"255.255.0.0"
Set the subnet mask:
SYST:COMM:LAN:SMAS "255.255.255.0"
SYST:COMM:LAN:UPD
l
If DHCP is enabled (SYSTem:COMMunicate:LAN:DHCP ), the specified subnet mask is not used. However, if the
DHCP server fails to assign a valid IP address, the instrument uses the AutoIP subnet mask.
l
A value of "0.0.0.0" or "255.255.255.255" indicates that subnetting is not being used.
l
This setting is non-volatile; it will not be changed by power cycling or
l
The subnet mask is set to "255.255.0.0" when the instrument is shipped from the factory or after a SYS-
Tem:SECurity:IMMediate command.
l
CURRent: read address currently being used by the instrument.
l
STATic: read static address from non-volatile memory. This address is used if DHCP is disabled or unavailable.
SYSTem:COMMunicate:LAN:TELNet:PROMpt "<
string
>"
SYSTem:COMMunicate:LAN:TELNet:PROMpt?
Sets the command prompt seen when communicating with the instrument via Telnet.
Parameter
Typical Return
String of up to 15 characters
"Command>"
Set the command prompt:
SYST:COMM:LAN:TELN:PROM "Command>"
l
A Telnet session is typically started from a host computer shell:
telnet <
IP_address
> <
port
>
For example: telnet 169.254.4.10 5024
To exit a Telnet session, press <
Ctrl-D
>.
312
Agilent 33500 Series Operating and Service Guide