Quick Start
resents the amount of time per cycle that the square wave is at a high level.
3.
Enter the desired duty cycle.
Using the numeric keypad or the knob and arrows, select a duty cycle value of 75. If you are using the numeric key-
pad, press the Percent softkey to finish the entry. The instrument adjusts the duty cycle immediately and outputs
a square wave with the specified value (if the output is enabled).
Configure a Pulse Waveform
You can configure the instrument to output a pulse waveform with variable pulse width and edge time. The following
steps configure a 500 ms periodic pulse waveform with a pulse width of 10 ms and edge times of 50 ns.
1.
Select the pulse function.
Press the
Waveforms
key and choose
Pulse
to select the pulse function.
2.
Set the pulse period.
Press the
Units
key and then press
Frequency/Period
to choose
Period
. Then press
Parameters
and choose
Period
. Set the period to 500 ms.
3.
Set the pulse width.
Press
Parameters
and then
Pulse Width
. Then set the pulse width to 10 ms. The pulse width represents the
24
Agilent 33500 Series Operating and Service Guide