Triggering
Front-Panel:
SCPI:
Manual Triggering
Manual trigger mode (front panel only): you manually trigger the instrument by pressing
Trigger
. The instrument ini-
tiates one sweep or burst for each time you press
Trigger
. The button is lit when you are in the trigger menu and the
instrument is waiting for a manual trigger. The button blinks when the instrument is waiting for a manual trigger, but
you are not in the trigger menu. The key is disabled when the instrument is in remote.
External Triggering
In external trigger mode, the instrument accepts a hardware trigger at the rear-panel
Ext Trig
connector. The instru-
ment initiates one sweep or burst each time Ext Trig receives a TTL pulse with the specified edge. The external trigger
mode is like the manual trigger mode except that you apply the trigger to the
Ext Trig
connector.
See
, below.
Front-Panel:
To specify whether the instrument triggers on a rising or falling edge, press Trigger Setup and select the edge direction
by pressing Slope.
SCPI:
TRIGger:SLOPe {POSitive|NEGative}
Software (Bus) Triggering
Available only from remote interface, this is similar to manual trigger mode from the front panel, but you trigger the
instrument with a bus trigger command. The instrument initiates one sweep or outputs one burst each time a bus
trigger command is received. The key blinks when a bus trigger command is received.
To select the bus trigger source, send
.
To trigger instrument from remote interface (GPIB, USB, or LAN) when Bus source is selected, send
or
(trigger). The front-panel
Trigger
key is illuminated when the instrument is waiting for a bus trigger.
Timer Triggering
The timer trigger mode issues triggers a fixed period apart. To select the bus trigger source, send
104
Agilent 33500 Series Operating and Service Guide