This section describes the instrument's calibration security system.
Security Overview
A security code prevents accidental or unauthorized instrument adjustments.
l
For models 33521A and 33522A, the security code is set to
AT33520A
when instrument ships from factory. For
all other models, the security code is set to
AT33500
.
l
Once you enter a security code, that code must be used for both front-panel and remote operation. If you secure
the instrument from the front panel, you must use that same code to unsecure it from the remote interface.
l
This setting is non-volatile; it will not be changed by power cycling or
l
Security code rules:
Unquoted string up to 12 characters
Must start with letter (A-Z)
May contain letters, numbers (0-9) and underscores
l
Front Panel:
l
SCPI:
Unsecure Instrument Without Security Code
See
Electrostatic Discharge (ESD) Precautions
before beginning this procedure.
1. Disconnect power cord and all input connections.
2. Disassemble the instrument (see
).
374
Agilent 33500 Series Operating and Service Guide