AC Amplitude (high-impedance) Adjustment
Setup
Nominal Signal
DC Level
22
+0.68 V
Output of -24 dB range
23*
-0.68 V
Output of -24 dB range
24
+1.7 V
Output of -16 dB range
25*
-1.7 V
Output of -16 dB range
26
+4.3 V
Output of -8 dB range
27*
-4.3 V
Output of -8 dB range
28
+10.8 V
Output of 0 dB range
29*
-10.8 V
Output of 0 dB range
30
+0.044V
Output of -48 dB High DC range
31*
-0.044V
Output of -48 dB High DC range
32
+0.11 V
Output of -40 dB High DC range
33*
-0.11 V
Output of -40 dB High DC range
34
+0.28 V
Output of -32 dB High DC range
35*
-0.28 V
Output of -32 dB High DC range
36
+0.68 V
Output of -24 dB High DC range
37*
-0.68 V
Output of -24 dB High DC range
* Constants are stored after completing this setup.
3. Using the numeric keypad or knob, adjust the displayed voltage at each setup to match the measured voltage.
Select
ENTER VALUE
.
4. After performing setup 37:
a. If your calibration procedures require you to verify this adjustment, exit the calibration menu and perform
Amplitude (high-impedance) Verification.
b. If you are making all of the adjustments and then verifying the instrument’s performance, continue with the
next procedure in this section.
394
Agilent 33500 Series Operating and Service Guide