-8 dB Range Flatness Verification
Use
Waveform Generator
Measurement
50 Ω
Sine
1.200
Vrms
1.000 MHz
100%
±
1.74%
0 dB
± 0.15
dB
Q
50 Ω
Sine
1.200
Vrms
2.000 MHz
100%
±
1.74%
0 dB
± 0.15
dB
50 Ω
Sine
1.200
Vrms
5.000 MHz
100%
±
1.74%
0 dB
± 0.15
dB
50 Ω
Sine
1.200
Vrms
10.00 MHz
100%
±
3.51%
0 dB
± 0.30
dB
50 Ω
Sine
1.200
Vrms
15.00 MHz
100%
±
3.51%
0 dB
± 0.30
dB
50 Ω
Sine
1.200
Vrms
20.00 MHz
100%
±
3.51%
0 dB
± 0.30
dB
50 Ω
Sine
1.200
Vrms
22.00 MHz
(unnecessary for 20 MHz
instruments)
100%
±
4.71%
0 dB
± 0.40
dB
50 Ω
Sine
1.200
Vrms
27.00 MHz
(unnecessary for 20 MHz
instruments)
100%
±
4.71%
0 dB
± 0.40
dB
50 Ω
Sine
1.200
Vrms
30.00 MHz
(unnecessary for 20 MHz
instruments)
100%
±
4.71%
0 dB
± 0.40
dB
6. Compare the measured value to the test limits shown in the table.
7. Two-channel instruments only: Connect the AC voltmeter to channel 2 and repeat steps 2 through 6.
382
Agilent 33500 Series Operating and Service Guide