Output Configuration
To select another waveform:
l
For example, to specify a DC signal:
To produce the DC output:
l
SCPI:
FUNCtion {SIN|SQU|RAMP|PULSe|NOIS|DC|PRBS|ARB}
The
command configures a waveform with one command.
Output Frequency
As shown below, the output frequency range depends on the function (default frequency 1 kHz for all functions).
Function
Minimum Frequency
Maximum Frequency
Sine
1 µHz
30 MHz
Square
1 µHz
30 MHz
Ramp/Tri.
1 µHz
200 kHz
Pulse
1 µHz
30 MHz
PRBS
1 µbps
50 Mbps
Arbitrary
1 µSa/s
250 MSa/s
Agilent 33500 Series Operating and Service Guide
57