MARKer Subsystem
MARKer Subsystem
The MARKer subsystem configures the point within an arbitrary waveform, sweep, or burst at which the front-panel
Sync
signal goes low.
Commands and Queries
l
- cycle of a burst at which
Sync
signal goes low
l
- frequency at which
Sync
signal goes low
l
- point in an arbitrary waveform at which
Sync
signal goes low
Each of these commands causes sync/marker to transition to high at start of burst, sweep, or arbitrary waveform (
may reverse this.).
[SOURce[1|2]:]MARKer:CYCLe {<
cycle_num
>|MINimum|MAXimum}
[SOURce[1|2]:]MARKer:CYCLe? [{MINimum|MAXimum}]
Sets the marker cycle number at which the front-panel
Sync
signal goes low in a burst mode operation.
may reverse this..
Parameter
Typical Return
Whole number from 2 to number of cycles in the burst plus one (1),
default 2
+2.00000000003
Set the marker cycle to 2000:
MARK:CYCL 2000
l
This is valid only if burst is enabled and
is MARKer.
l
With burst enabled, marker cycle must be less than number of cycles in burst plus one. Attempting to set the
marker cycle outside this range will set marker cycle equal to middle of burst. From the remote interface, a "Set-
tings conflict" error will also be generated.
[SOURce[1|2]:]MARKer:FREQuency {<
frequency
>|MINimum|MAXimum}
[SOURce[1|2]:]MARKer:FREQuency? [{MINimum|MAXimum}]
Sets the marker frequency at which the front-panel
Sync
signal goes low during a sweep.
may
reverse this..
Parameter
Typical Return
Any frequency between start and stop frequency, default 500 Hz
+2.00000000003
Set marker frequency to 2 kHz:
MARK:FREQ 2000
l
This is valid only if burst is enabled and
is MARKer.
244