Disassembly
board toward front of instrument to disengage tabs on power supply cover. Lift main board out.
4. Remove Power Supply. Disconnect input power to power supply board (blue and brown wires). Disconnect green
ground connector on power supply board. Remove screw securing power supply cover to chassis. Slide power
supply assembly toward front of instrument and remove.
Always re-attach green ground wire to power supply before operating instrument.
5. The remaining assemblies can be removed from the chassis if needed.
Front Panel Disassembly
1. Pull the knob straight off. Remove six T8 screws securing front panel bracket to front panel assembly. Lift out
bracket.
Agilent 33500 Series Operating and Service Guide
425