CALibration Subsystem
CALibration Subsystem
The CALibration subsystem is used to calibrate the instrument.
Commands and Queries
CALibration:SECure:CODE <new_code>
CALibration:SECure:STATe {OFF | ON} [,<code>]
CALibration[:ALL]?
Performs a calibration using the calibration value (
). The instrument must be unlocked (
) to calibrate.
Parameter
Typical Return
(none)
+0 (pass) or +1 (fail)
Calibrate using the current value:
CAL?
l
should always precede the CALibration? query.
l
Increments the instrument’s calibration count (
).
l
Modifies the volatile version of the calibration constants. Use
to save these constants in non-
volatile memory at end of calibration.
CALibration:COUNt?
Returns the number of calibrations performed. Read and record the initial count when you receive your instrument
from the factory.
Parameter
Typical Return
(none)
+117
180