OUTPut Subsystem
Parameter
Typical Return
Set output sync mode to CARRier:
OUTPut:SYNC:MODE CARR
l
The following table details the command's behavior:
<mode>
Sync Behavior
Conditions
NORMal
Sync follows envelope of burst sig-
nal.
When BURSt is on.
Sync follows envelope of sweep sig-
nal.
When SWEep is on.
Sync follows modulating signal.
When modulating and modulation source is internal.
Sync follows FUNC signal.
All other conditions
CARRier
Sync follows current
SOURce:MARKer:POINt setting.
When BURSt or SWEep is on and FUNC is ARB.
Sync follows FUNC signal while burst
is on.
When BURSt is on and FUNC is not ARB.
Sync follows FUNC signal.
All other conditions
MARKer
Sync follows current
SOURce:MARKer:POINt setting.
When in CW mode and FUNC is ARB.
When modulating, modulation source is internal, and FUNC is
ARB or <
modulation
>:INT:FUNC is ARB.
When modulating, modulation source is external, and FUNC is
ARB.
Sync follows current
SOURce:MARKer:CYCLe setting.
When BURST is on.
Sync follows current
SOURce:MARKer:FREQuency set-
ting.
When SWEEP is on.
Sync follows modulating signal.
When modulating, modulation source is external, FUNC is not
ARB, and INT:FUNC is not ARB.
Sync follows FUNC signal.
All other conditions
OUTPut[1|2]:SYNC:POLarity {NORMal|INVerted}
OUTPut[1|2]:SYNC:POLarity?
Sets the desired output polarity of the
Sync
output to trigger external equipment that may require falling or rising
edge triggers.
262