General Maintenance
10
Troubleshooting and Maintenance Guide
151
To Reinstall a Filament
Materials needed
•
Filament assembly (G2590-60053)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
Procedure
1
Install the new filament (
).
2
Remove the screw holding the filament to the ion source
body.
3
Slide off the filament assembly.
4
Orient the assembly so that the filament is next to the ion
source body.
5
Replace the screw to the ion source body.
6
After installing the filament, verify that it is not grounded to
source body.
7
Reinstall the ion source. See the
8
Close the analyzer chamber.
9
Pump down the MSD.
10
Autotune the MSD.
11
In the
Edit Parameters
dialog box (
Instrument/Edit MS Tune
Parameters
), select the other filament.
12
Autotune the MSD again.
13
Select and use the filament that gives the best results.
If you decide to use the first filament, run Autotune again to
make sure the tune parameters are compatible with the
filament.
14
Select
Save Tune Parameters
from the
File
menu.
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...