Troubleshooting and Maintenance Guide
11
Maintaining the Split/Splitless Inlet
To Replace the Filter in the Split Vent Line for the Split/Splitless Inlet
1
Gather the following:
•
“Consumables and Parts for the
•
T-20 Torx screwdriver
2
Load the maintenance method and wait for the GC to
become ready.
3
Remove the pneumatics cover on the top of the instrument.
4
Loosen the screws on the retaining clip.
WA R N I N G
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands
.
WA R N I N G
The split vent trap may contain residual amounts of any samples
or other chemicals you have injected into the GC. Follow your
company’s safety procedures for handling these types of
substances while replacing the trap filter cartridge.
Split vent valve
Split vent trap
Retaining clip
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...