Troubleshooting and Maintenance Guide
11
Maintaining the Split/Splitless Inlet
To Replace the Gold Seal on the Split/Splitless Inlet
1
Gather the following:
•
•
Replacement washer
•
1/4-inch wrench (for column)
•
1/2-inch wrench
•
Lint-free gloves
2
Load the maintenance method and wait for the GC to
become ready.
3
Open the Guard Column enclosure door.
4
Remove the inlet guard column from the inlet. Cap the open
end of the column to prevent contamination.
WA R N I N G
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands
.
Figure 22
Inlet
Inlet