Troubleshooting Tasks
9
Troubleshooting and Maintenance Guide
107
•
The split vent has a 1/8-in Swagelok threaded fitting. Create
and use a 1/8-in tub adapter (as shown above) to covert the
1.8-in threaded fitting into a 1/8-tube. This prevents the
rubber flowmeter tubing from leaking around the threads,
which will result in leakage and thus an incorrect flow
reading.
•
The septum purge is a 1/8-in tube. Use the red rubber
adapter shown to measure flows.