General Maintenance
10
Troubleshooting and Maintenance Guide
139
To Remove the EI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
Procedure
1
Vent the instrument.
2
Open the analyzer chamber. See the
5975T LTM GC/MSD
Operation Manual.
3
Make sure you use an antistatic wrist strap and take other
antistatic precautions before touching analyzer components.
4
Disconnect the seven wires from the ion source. Do not bend
the wires any more than necessary (
5
Trace the wires for the ion source heater and temperature
sensor to the feedthrough board. Disconnect them there.
6
Remove the thumbscrews that hold the ion source in place.
7
Pull the ion source out of the source radiator.
Table 7
Ion source wires
Wire color
Connects to
Number of leads
Blue
Entrance lens
1
Orange
Ion focus
1
White
Filament 1 (top
filament)
2
Red
Repeller
1
Black
Filament 2
(bottom
filament)
2
C A U T I O N
Pull on the connectors, not on the wires.
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...