144
Troubleshooting and Maintenance Guide
10
General Maintenance
Procedure
3
If the contamination is serious, such as an oil backflow into
the analyzer, seriously consider replacing the contaminated
parts.
Figure 13
Source parts to be cleaned
Source body
Repeller
Interface socket
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens
C A U T I O N
The filaments, source heater assembly, and insulators cannot be
cleaned ultrasonically. Replace these components if major
contamination occurs.