General Maintenance
To Attach a Capillary Column Using SilTite Metal Fittings
This procedure is used to attach a capillary column to the
Ultimate Union.
1
Gather the following:
•
SilTite ferrules (see
)
•
Swaging nut for SilTite ferrules (G2855-20555)
•
Two 1/4-inch open-end wrenches
•
One 7/16-inch open-end wrench
•
Column cutting tool (5181-8836)
•
Internal nut (G2855-20530)
•
Lint free gloves
2
Pass the tubing end through the internal nut and SilTite
ferrule leaving approximately 1 cm of fused silica tubing
protruding beyond the ferrule. Thread the swaging nut onto
the internal nut with the tube protruding.
Table 6
Available SilTite metal ferrule packages
Part number
SilTite ferrule description
5188-5361
For 0.2- to 0.25-mm columns
5188-5362
For 0.32-mm columns
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...