Maintaining the Split/Splitless Inlet
11
Troubleshooting and Maintenance Guide
177
5
Completely loosen the two screws that secure the split vent
valve in place.
6
Lift the filter trap assembly and split vent valve from the
mounting bracket together and unscrew the split vent front
weldment on the filter trap assembly. Be careful not to
stress the tubing between the split vent valve and the trap.
7
Remove the old filter cartridge and two O-rings.
8
Verify the new O-rings are seated properly on the new filter
cartridge.
9
Install the new filter cartridge then reassemble the trap. Do
not fully tighten yet.
10
Place the filter trap assembly in the mounting bracket and
install the retaining clip.
11
Install the split vent valve.
Split vent trap
O-ring
Cartridge
O-ring