Not Ready Symptoms
The Inlet Pressure or Flow Fluctuates
A fluctuation in inlet pressure causes variations in the flow rate
and retention times during a run. Do the following:
•
Check if the gas purifier or gas generator is operating at or
near capacity.
•
Check the supply gas for sufficient delivery pressure.
•
Verify that the supply pressure regulator is functioning
properly.
•
Check for leaks. (See
•
Check for large restrictions in the inlet liner or split vent
trap.
•
Verify that the correct liner is installed.
•
Check for a restriction in headspace, purge and trap, and
any other external sampling devices.