106
Troubleshooting and Maintenance Guide
9
Troubleshooting Tasks
To Measure a Split Vent or Septum Purge Flow
Note that the GC reports flows calibrated to 25 °C and
1 atmosphere. Correct flowmeter results accordingly.
Septum purge and split vent flows exit through the pneumatic
module at the top rear of the instrument. See
To measure split vent or septum purge flows, attach the
flowmeter to the appropriate tube.
WA R N I N G
Hydrogen (H
2
) is flammable and is an explosion hazard when
mixed with air in an enclosed space (for example, a flow meter).
Purge flowmeters with inert gas as needed. Always measure
gases individually.
Figure 4
Top rear view of the instrument
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...