To Refill the EI Calibration Vial
Materials needed
•
PFTBA (05971-60571)
Procedure
1
Stop any tuning or data acquisition.
2
Turn off the analyzer. There are several ways to do this:
•
In the
Tune and Vacuum Control
view, select
MS OFF
from the
Execute
menu.
•
In the
Instrument Control
view in the
Edit Parameters
dialog
box, select
MS OFF
from the
Execute
menu.
3
Remove the analyzer window cover.
4
Loosen the calibration vial collar (
). Do not remove
the collar.
5
Pull the calibration vial out. You may feel some resistance
due to residual vacuum.
6
Syringe or pipette PFTBA into the vial. With the vial
vertical, the liquid should be just below the end of the
internal tube, approximately 70 µL of sample.
7
Push the calibration vial into the valve as far as possible.
8
Withdraw the vial 1 mm. This prevents damage when you
tighten the collar.
9
Turn the collar clockwise to tighten it.
Figure 7
Removing the EI calibration vial
Calibration vial
Collar
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...