![Agilent Technologies 5975T LTM GC/MSD Troubleshooting And Maintenance Manual Download Page 124](http://html1.mh-extra.com/html/agilent-technologies/5975t-ltm-gc-msd/5975t-ltm-gc-msd_troubleshooting-and-maintenance-manual_2867607124.webp)
124
Troubleshooting and Maintenance Guide
10
General Maintenance
11
Place the new main 0-ring into the groove in the pump
frame, ensuring first that the groove is clean.
12
Replace the outboard housing, ensuring that the tip seal has
not fallen out of its groove. See
13
Replace the four M5 bolts in the outboard housing. See
.
14
Reconnect the electrical connector at the front cowling.
15
Replace the front cowling and the four M5 bolts that hold it
in place. See
16
Reconnect the pump to electrical power.
17
Reposition the foreline pump, providing slack in the tubing
and cables.
The foreline pump may be located on the floor or on the lab
bench next to the GC/MSD.
18
Pump down the instrument.
Figure 6
Changing the dry pump tip seal
Front cowling bolts
Front cowling
Frame bolts
Outboard housing
Orbiting scroll
Tip seals and main o-ring are not shown here
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...