Maintaining the Split/Splitless Inlet
11
Troubleshooting and Maintenance Guide
175
5
Loosen and remove the reducing nut. Remove the washer
and seal inside the reducing nut.
6
Put on gloves to protect the new gold seal and washer from
contamination. Put a new washer in the reducing nut and
place the new gold seal on top of it (raised portion facing
down).
7
Replace the reducing nut and tighten securely with a
wrench.
8
Install the inlet guard column.
9
Bakeout contaminants. (See the
5975T LTM GC/MSD
Operation Manual
.)
10
Restore the analytical method.
11
Reset the EMF counter.
12
Check for leaks.
C A U T I O N
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Inlet gold seal
Washer
Reducing nut
Gold seal side view
Make sure raised portion faces
down.
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...