136
Troubleshooting and Maintenance Guide
10
General Maintenance
6
Reinstall the vent valve knob.
7
Pump down the instrument.
Figure 10
Vent valve O-ring
Vent valve knob
Vent valve O-ring
C A U T I O N
Be very careful when reinstalling the vent valve knob. It is possible
to cross thread the knob and damage the threads in the valve body.
Be sure the O-ring stays in place.