General Maintenance
10
Troubleshooting and Maintenance Guide
141
To Disassemble the EI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Wrench, open-end, 10 mm (8710-2353)
Procedure
1
Remove the ion source. See the
2
Remove the filaments (
).
3
Separate the repeller assembly from the source body. The
repeller assembly includes the source heater assembly,
repeller, and related parts.
4
Remove the repeller.
5
Unscrew the interface socket. A 10-mm open-end wrench
fits the flats on the interface socket.
6
Remove the setscrew for the lenses.
7
Push the lenses out of the source body.
N O T E
Video shows the standard Ion Source Assembly procedures.
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...