74
Troubleshooting and Maintenance Guide
4
Not Ready Symptoms
Instrument Never Becomes Ready
Normally the instrument becomes ready after flows and
temperatures reach setpoint. If the instrument does not become
ready after a long period of time, perform the following using
the LCP:
•
Press
[Status]
or a component key (for example,
[Front inlet]
)
to see which setpoints or conditions are not ready.
•
Check for a sampler problem.
•
Check for a data system problem.
•
If performing manual injections in splitless or gas-saver
mode, you may need to press
[Prep Run]
to prepare the inlet
for the injection. Do this, for example:
•
To toggle the inlet purge valve before a splitless injection
•
To prepare for a pulsed injection
•
To turn off gas saver.
For more information on
[Prep Run]
, see the
5975T LTM
GC/MSD Operation Manual
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...