Electronics
14
Troubleshooting and Maintenance Guide
225
Event connector
This connector provides two passive contact closures and two
24-volt outputs for controlling external devices. The outputs are
controlled by valve drivers 5 through 8.
BCD connector
This connector provides two control relays and a BCD input for
a stream selection valve.
Sampler connector
This connector provides an injector, usually the front injector.
(For 7693A, the configuration does not matter.)
C A U T I O N
This connector is similar to the Event connector. Plugging a non
BCD cable into the BCD connector can damage the GC.