General Maintenance
To Replace the Electron Multiplier Horn
Materials needed
•
Electron multiplier horn (G3170-80103)
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
Procedure
1
Vent the instrument.
2
Open the analyzer chamber.
3
Open the retaining clip (
). Lift the arm of the clip
up and then swing the clip away from the electron multiplier
horn.
4
Remove the electron multiplier horn.
5
Install the new electron multiplier horn.
6
Close the retaining clip.
The signal pin on the horn must rest
on the outside
of the
loop in the contact strip.
Do not
put the signal pin on the
inside of the loop in the contact strip. Incorrect installation
will result in poor sensitivity or no signal.
7
Close the analyzer chamber.
Summary of Contents for 5975T LTM GC/MSD
Page 1: ...Agilent Technologies Agilent 5975T LTM GC MSD Troubleshooting and Maintenance Guide...
Page 10: ...10 Troubleshooting and Maintenance Guide...
Page 32: ...32 Troubleshooting and Maintenance Guide 2 Concepts and General Tasks...
Page 72: ...72 Troubleshooting and Maintenance Guide 3 Chromatographic Symptoms...
Page 86: ...86 Troubleshooting and Maintenance Guide 5 ALS Symptoms...
Page 98: ...98 Troubleshooting and Maintenance Guide 7 Instrument Power On and Communication Symptoms...
Page 108: ...108 Troubleshooting and Maintenance Guide 9 Troubleshooting Tasks...
Page 166: ...166 Troubleshooting and Maintenance Guide 10 General Maintenance...
Page 214: ...214 Troubleshooting and Maintenance Guide 13 Analyzer...