General Maintenance
10
Troubleshooting and Maintenance Guide
149
To Remove a Filament
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5-mm (8710-1570)
Procedure
1
Vent the MSD.
2
Open the analyzer chamber.
3
Remove the ion source. See the
4
Remove the filament(s) to be replaced (
).
Figure 15
Installing the EI ion source
Ion source
Thumbscrews
Source radiator
WA R N I N G
The analyzer operates at high temperatures. Do not touch any part
until you are sure it is cool.